Fluke Calibration 9143-C-P-156

$4,247.00

In stock

SKU: NSKU1623 Category: Tag:
  • Field Metrology Well, 33 °C to 350 °C, w/9143-INSX, w/Process Electronics
  • Temperature Range at 23 °C: 33 °C to 350 °C (91 °F to 662 °F)
  • Display Accuracy: ± 0.2 °C Full Range
  • Stability: ± 0.02 °C at 33 °C ± 0.02 °C at 200 °C ± 0.03 °C at 350 °C
  • Axial Uniformity at 40 mm (1.6 in): ± 0.04 °C at 33 °C ± 0.1 °C at 200 °C ± 0.2 °C at 350 °C
  • Radial Uniformity: ± 0.01 °C at 33 °C ± 0.015 °C at 200 °C ± 0.02 °C at 350 °C
  • Loading Effect (with a 6.35 mm reference probe and three 6.35 mm probes): ± 0.015 °C Full Range
  • Hysteresis: 0.03°C
  • Operating Conditions: 0 °C to 50 °C, 0 % to 90 % RH (non-condensing)
  • Environmental Conditions (for all specifications except temperature range): 13 °C to 33 °C
  • Immersion (Well) Depth: 150 mm (5.9 in)
  • Insert OD: 25.3mm (1.00 in)
  • Heating Time: 5 min: 33 °C to 350 °C
  • Cooling Time: 32 min: 350 °C to 33 °C 14 min: 350 °C to 100 °C
  • Resolution: 0.01 °
  • Display: LCD, °C or °F user-selectable
  • Size (H x W x D): 290 mm x 185 mm x 295 mm (11.4 x 7.3 x 11.6 in)
  • Weight: 7.3 kg (16 lb)
  • The process version provides a built-in two-channel thermometer readout that measures PRTs, RTDs, thermocouples, and 4-20 mA transmitters which includes the 24 V loop supply to power the transmitter.

>

Each Unit Includes

  • Insert C
  • RS-232 and 9930 Interface-it control software
  • Test Leads
Average Rating

4.00

5 Star
0%
4 Star
0%
3 Star
0%
2 Star
0%
1 Star
0%
Add a review

Your email address will not be published. Required fields are marked *

Reviews

  1. by Victor Zhang

Returns Policy

We accept returns within 30 days of delivery. To qualify, items must be unused, in their original condition, and returned with all original packaging and accessories.

Once we receive and inspect the returned item, we will process your refund within 7 business days. Refunds will be issued to the original payment method.

Please note: Shipping costs are non-refundable unless the return is due to a defective or incorrect item.

Ask a Question




Main Menu