No products in the cart.
- Yokogawa Model # CL135
- Approved for comformity to safety standard EN61010-1, EN61010-2-031, EN61010-2-032 (CAT. III 600 V)
- Method of detection: True RMS
- Display: LCD (Digital display: 1999 counts)
- Response Time: Approx. 1 second (approx. 2 seconds on resistance renge)
- Range Switching: Manual-range
- Data Hold: On all range
- Ambient temperature and humidity: -10 to 50 °C (no condensation), up to 30 deg.C, 90% RH up to 40 deg.C, 75% RH, up to 50 deg.C, 45% RH
- Influence of external magnetic field: 2A or less at 400A/m
- Influence of conductor position: ±2% or less
- Circuit Voltage: 600Vrms or less
- Withstanding Voltage: 5.55kV AC for one minute
- Power Supply: 6F22(006P)9V ×1 or 6LR61 ×2
- Battery Life: Approx. 200 hours (continuous)
- Consumed Current: Approx. 2mA
- Auto power-off: Approx. 10 minutes
- Diameter of measurable conductor: 30mm at maximum
- Dimensions: Approx. 93(W) × 210(H) × 40(D)mm
- Weight: Approx. 400g
>
Each Unit Includes
- Users manual (IM CL130)
- Batteries
- Carrying case (93032)
Returns Policy
We accept returns within 30 days of delivery. To qualify, items must be unused, in their original condition, and returned with all original packaging and accessories.
Once we receive and inspect the returned item, we will process your refund within 7 business days. Refunds will be issued to the original payment method.
Please note: Shipping costs are non-refundable unless the return is due to a defective or incorrect item.
Reviews
There are no reviews yet.